GE Senographe 2000 D Operator's Manual page 135

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Note:
Query
Option
Detach
Preview option
2-2-2
The Preview of CD–R content option automatically displays a list of the data to be recorded on the
CD-R after clicking on the (OK) button or after using "drag-and-drop" to start recording, but before
saving starts. The estimated time needed to write the data to the CD-R is also shown.
At this stage, you can still cancel the save operation and modify your selection if required.
If you have not set the Preview of CD–R content option, saving starts immediately after clicking on the
(Save) button, or after using "drag-and-drop".
D To set this option on or off, click on the (CD-R) icon in the Network/Archive panel
on the main Browser and select [Option] in the drop-down menu.
Query
Option
Detach
Checking the CD-R
2-2-3
You are recommended to check for successful completion by browsing the CD-R at the end of the
save operation.
D To do so, click the (CD-R) icon on the main Browser, then select [Query] in the drop-down menu. If
all the patients selected for the save operation are not displayed on the Media Browser, an error
has occurred while recording the CD, and you will have to repeat the save operation.
D An error message appears next to the patient name if the associated information has not been
saved. If this occurs:
– Make a list of the patients correctly saved and those with an error.
– Insert a new CD-R and repeat the save process.
– If there are still errors, consult GE Service.
Availability of system while writing to CD-R
2-2-4
No other operation can be performed on the system while saving images on a CD-R. For a full CD-R
the save operation can take up to 45 minutes. If the system is required for image acquisition, the save
operation can be abandoned:
D Enter the Worklist in the usual way and request a new exam.
D A message is displayed asking if the write operation should be aborted. Answer (Yes) or (Cancel).
D If you choose (Yes), the write operation is abandoned, and the AWS application restarts. You must
wait for completion of the restart process and then repeat the request for the new exam. Note that
the partially written CD-R is not useable.
While selecting items (before you start the save), the operation can be
abandoned by clicking on the [CD-R] icon and selecting [Detach] in the
drop-down menu. This stops the operation. The CD is ejected automatically
and can be used again.
Senographe 2000 D Acquisition System
133
OM 5179217–1–100