TECHNICAL SPECIFICATIONS
1-1
Electrical Specification
Mode of operation: Continuous operation with intermittent loading.
Line voltage specifications
1-1-1
D Single-phase input voltage (phase-neutral or phase-phase):
– 200/208/220/240/380/415 V (+ 10 %)
D Can be powered by single-phase 440 V (+ 10%) using optional transformer.
D Line frequency specifications
D Protection against electrical shocks: class 1, type B.
D The maximum line current corresponds to the use of the technique factors 30 kV, Mo track, large
focal spot, and 100 mAs.
kVA load characteristics
1-1-2
• Maximum power in standby: 1.5 kVA.
• Maximum instantaneous power (during exposures, up to 6 seconds) 9 kVA.
• Power factor: 0.6
• Line current crest factor: 1.7 at 200 V to 2 at 415 V.
Input impedance
1-1-3
The apparent resistance of the mains supply R
drop of 6% at the maximum power load of 9 kVA.
For a nominal supply voltage of 380 V, the line resistance R
Ω
R
< 1
L(380)
For other supply voltages, calculate the equivalent permissable resistance value in ohms as follows:
R
= RL(380) x (U/380)
L(U)
Where:
U = local input voltage
R
= total apparent resistance (two–wire) of the supply at the local voltage (U).
L(U)
R
= total apparent resistance (two–wire) of the supply at 380 V.
L(380)
Thus, as R
must be less than 1
L(380)
Generator Output (excluding tube)
1-1-4
D 22 thru 49 kV,
D 20 thru 130 mA.
Duty Cycle
1-1-5
The generator is always limited by the tube:
CHAPTER 15 – SPECIFICATIONS
50 or 60 Hz (+ 5%)
must be less than that which would cause a voltage
L
Ω,
the value of RL(U) must be less than (U/380).
165
Senographe 2000 D Acquisition System
must be less than 1
L(380)
OM 5179217–1–100
Ω: