GE Senographe 2000 D Operator's Manual page 163

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message on AWS screen
AA
"Shutdown Process started
10
..."
System is down.
AA
11
System will power off within 30
seconds: do NOT reboot now!
(if system does not turn off,
switch off UPS manually)
AA
"System not ready, exam
12
forbidden"
AA
UPS battery needs replacing.
13
Please call service
UPS output overload.
AA
14
Please call service
BB These messages relate to boot/reset events:
"AWS – IDC communication
BB
1
failure, exam forbidden."
AWS/SENO communication
BB
2
failure, exam forbidden
BB
AWS not functional, exam
3
forbidden
Senographe 2000 D Acquisition System
Explanation
The AWS shutdown process has
been started because a UPS low
battery condition was detected
(battery power must be available
to guarantee safe operation and
shutdown in the event of power
failure).
The message is displayed at the
end of the normal shutdown
sequence, when the UPS is
about to shut down, to prevent
users from trying to reboot.
A system component is not yet
ready (after system stsrt-up).
The UPS selftest has detected
that the battery requires
replacement.
An UPS overload condition has
been detected.
Communication between the
AWS and the acquisition system
was not established at boot time,
or has been lost during
application.
Communication between the
AWS and the Gantry system was
not established at boot time, or
has been lost during application.
One of the system components
did not boot (or did not reset)
correctly.
161
OM 5179217–1–100
Corrective action
1. Stop system usage and wait for the end
of the shutdown sequence.
2. Check the mains power supply.
3. Check (on UPS panel) that the batteries
are charging. Do not attempt to restart the
system until the batteries have recharged.
4. If the problem persists, contact GEMS
Field Service.
No action is normally required.
If the system fails to shut down completely
(the AWS screen continues to display the
message after one or two minutes), switch
it off by pressing the UPS OFF button at
the front of the Generator Cabinet.
Wait until system initialization is complete
before requesting an acquisition.
Stop system usage and contact GEMS
Field Service.
Stop system usage and contact GEMS
Field Service.
1. Click OK on the pop-up window
2. Restart the Browser.
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.
1. Click OK on the pop-up window
2. Reset the gantry (power OFF/ON),
and/or restart the Browser.
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.
1. Click OK on the pop-up window
2. Restart the Browser.
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.