GE Senographe 2000 D Operator's Manual page 155

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
I78
GAN
Blown Fuse
I79
GAN
Failure
I85
GAN
Force < –4
I86
GAN
Decompres.
I90
GAN
None
I98
GAN
Column
Fault
I99
GAN
Column
Fault
Message on
Explanation
Control Console
Blown Fuse
A fuse has failed.
Power Supply
A power supply has
Failure
Force < – 4 kgf
Decompression
Compression paddle
hit an obstacle during
up–movement
at start–up.
Compression not
Auto–Decompres
The automatic
sion is Off
decompression mode
has been inhibited by
Column Up
Impossible
up–movement is not
Column Down
Impossible
down–movement is
not possible.
153
Senographe 2000 D Acquisition System
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Switch OFF Gantry, then switch ON.
failed.
If problem continues, note ALL
messages and contact GEMS Field
Service.
Use the decompression pedal to
raise the paddle to park position.
Compression should now be
possible.
possible.
If auto–decompress mode is wanted,
switch modes by changing the
parameter in the Setup/Medical
the user.
menu. Otherwise no action is
required; this information is normally
displayed.
Elevator
1. Check if any movement switches
(up/down, rotation) were pressed
possible.
during power ON. If yes, release the
necessary switches and restart the
Gantry system.
2. Check if obstacle preventing
up–movement.
3. Command down–movement.
4. Command up–movement.
5. If the problem continues, note ALL
messages and contact GEMS Field
Service.
Elevator
1. Check if any movement switches
(up/down, rotation) were pressed
during power ON. If yes, release the
necessary switches and restart the
Gantry system.
2. Check if obstacle preventing
down–movement.
3. Command up–movement.
4. Command down–movement.
5. If the problem continues, note ALL
messages and contact GEMS Field
Service.
OM 5179217–1–100