GE Senographe 2000 D Operator's Manual page 145

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
1.
ERROR MESSAGES
Any operational anomalies found by the system during Senographe 2000 D operation are signalled by
Error Messages. These may be displayed on the Column Readout and on the Control Console or in a
pop-up box on the AWS screen. They may be simple information messages requiring no action, or they
may indicate error or fault conditions requiring action from the operator and/or GEMS Service
engineers.
Note that messages displayed on the Gantry begin with three identification characters, which are
followed by the message itself. On the Control Console, the message is followed by seven further
identification characters. Thus, for example, the Column Readout might display I76 FAILURE while the
Control Console displays I76 SENSOR FAILURE 158/011. If it necessary to call GEMS Field Service
about an error, you should quote the entire message, complete with the identification characters.
The tables below list error messages which may occur, with explanations and recommended actions.
Table 1 lists error messages displayed on the Gantry: control console and/or column.
Source of the message is mentionned:
GAN = issued from Gantry, GEN = from generator, AWS = from Acquisition workstation.
Table 2 lists error messages displayed on the AWS screen.
Table 1 – Error Messages displayed on the Gantry
Message
Msg.
on Column
Source
No.
Readout
E 01
GEN
None
E02
GEN
None
E03
GEN
None
E04
GEN
None
E05
GEN
None
CHAPTER 14 – ERROR MESSAGES
Message on
Control Console
Generator failure
Focus bias failure
Rotor failure
Arm fatal failure
Failure during
exposure
143
Senographe 2000 D Acquisition System
Explanation
Problem reported in
Switch OFF Gantry, then switch ON.
HV section of
If the problem continues, note ALL
generator.
messages and contact GEMS Field
Service.
Problem reported in
Switch OFF Gantry, then switch ON.
tube focus bias
If the problem continues, note ALL
circuits.
messages and contact GEMS Field
Service.
Problem reported in
Switch OFF Gantry, then switch ON.
anode starter.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
Generator detects
Switch OFF Gantry, then switch ON.
error in dialog with
If the problem continues, note ALL
arm.
messages and contact GEMS Field
Service.
Current exposure
Switch OFF Gantry, then switch ON.
aborted due to failure
If the problem continues, note ALL
in equipment.
messages and inform GEMS Field
Service.
OM 5179217–1–100
Corrective Action