GE Senographe 2000 D Operator's Manual page 150

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
E86
GAN
Chksum
Error
I05
GEN
None
I06
GEN
None
I07
GEN
None
I12
GEN
None
I13
GEN
None
Message on
Explanation
Control Console
Arm checksum
Content of saved
error
memory has been
modified. Calibration
may be false.
Compression is
Exposure aborted
Exposure aborted
by mAs level
because not enough
mAs available to
continue exposure.
Abort: Detect sig.
Exposure aborted
out range
because brightness
received by detector
are not in range to
continue exposure.
Selected filter not
Filter selected by the
allowed : Al
operator not allowed.
The Senograph
automatically selects
the next correct filter.
Maximum mAs =
Maximum available
mAs are equal to xxx.
IDC was not
Image Detection
ready
Control system was
not ready to perform
an image acquisition.
148
Senographe 2000 D Acquisition System
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
inhibited.
1. Press RESET.
2. Reposition the patient.
3. Start exam again
4. If the problem continues, note ALL
messages and contact GEMS Field
Service.
5. Nevertheless, you can continue to
use the system with manual
acquisition.
1. Press RESET.
2. Reposition the patient.
3. Start acquisition again
4. If the problem continues, note ALL
messages and contact GEMS Field
Service.
5. Nevertheless, you can continue to
use the system with manual
acquisition.
1. Press "Filter selection" to have the
correct filter selected.
Information message only.
(If mAs exceeded, cell aborts
exposure at 2 or 3 mAs if entire
exposure cannot be made.)
1. Press the prep button again.
2. Switch OFF the Gantry
3. Switch ON
4. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
5. Restart the system
6. If the problem persists, note ALL
messages and inform GEMS Field
Service.
OM 5179217–1–100
Corrective Action

Advertisement

Table of Contents
loading

Table of Contents