GE Healthcare
REV 1
Table 2 – Error messages displayed on the AWS monitor
Message on AWS screen
AA These messages relate to power on/off events:
Could not connect to the
AA
1
remote provider
AA
Do you really want to power
2
off the system: yes or no?
"End of remote connection in
AA
3
progress"
"Image detection failure,
AA
4
exam forbidden"
"Initialization not complete,
AA
5
please wait"
"No Ups Power Backup"
AA
6
"Senograph failure, exam
AA
7
forbidden"
AA
"Shutdown Process started
8
..."
"Shutdown Process started
AA
9
..."
Senographe 2000 D Acquisition System
Explanation
The specified remote host could
not be found. It may be
temporarily off-line, its address
may have changed or been
incorrectly entered, the
application on the remote
provider may not be running, or
there may be a network error.
This message is displayed as
part of the normal shutdown
sequence.
A GE Insite connection is being
closed.
An image detection subsystem is
not ready for acquisition after
start-up.
A system initialization process is
in progress.
A UPS error has been detected
or communication with the UPS
has been lost.
A Senographe (DMR) subsystem
is not ready for acquisition after
start-up.
The AWS shutdown process has
been started because a UPS
primary power failure longer than
10 seconds has been detected.
The AWS shutdown process has
been started because
communication with the UPS
was lost for more than 30
seconds.
160
OM 5179217–1–100
Corrective action
1. Try again after a short wait.
2. If the message is repeated, check that
the server address is correct and that the
remote application is running.
3. If the problem persists, contact GEMS
Field Service.
No corrective action required (click Yes or
No as required)
No action is required; you can now use the
system.
1. Wait to see if the problem clears.
2. If the problem persists, contact GEMS
Field Service.
Wait until system initialization is complete.
1. Stop system usage and shut down the
system.
2. Check the mains power supply.
3. Check (on UPS panel) that the batteries
are charging. Do not attempt to restart the
system unless the batteries have been
recharged.
3. Restart the system.
3. If the problem persists, contact GEMS
Field Service.
1. Wait to see if the problem clears.
2. If the problem persists, contact GEMS
Field Service.
1. Wait for the end of the shutdown
sequence.
2. Check mains power supply.
3. Restart the system when power OK.
4. If the problem persists, contact GEMS
Field Service.
1. Stop system usage and wait for the end
of the shutdown sequence.
2. Restart the system
3. If the problem persists, contact GEMS
Field Service.