Control Console
Exposure too
The exposure is too
long
long (this message
should appear only in
manual mode).
Compression
Breast thickness is
sensor fault
badly calculated in
It is still possible to
continue the exam in
manual mode.
Wait Until End of
Acquisition is inhibited
Digital Acquisit.
until the previous
image has been
completely recorded.
Exp. Abort, Wait
Exposure in progress
Image Acquisition
has been aborted by
the operator or the
equipement but the
system processed and
display the image that
has been generate.
Abort by HV
arcing. Wait
Image Acq.
Abort by IDC.
Wait Image
Acquisition
Incorrect X_Ray
The position of the
Incidence
tube carrying arm does
not allow exposure: –
Position not locked at
S81 Image
One of the image
detection not
detection coponent is
ready
buzy or its condition to
acquire images is not
yet reached.
156
Senographe 2000 D Acquisition System
Increase the kV value or reduce the
mAs value.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
AOP mode.
messages and contact GEMS Field
Service.
No action is required. The display of
this message is normal during image
transfer to the workstation. It will be
cleared when system is ready for the
next acquisition.
If the problem persists, switch OFF
gantry, then switch it ON.
1. Look at the message displayed
when the reset button is highlighted
to know the real problem
2. Press reset and perform the action
required by the last displayed
message.
Correct the position of the tube
carrying arm.
0 degrees.
1. Wait for completion of system
setting.
2. If this X–ray inhibition stays
displayed more that 10 mn Switch
OFF the whole system and wait for
the end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL
messages and inform GEMS Field
Service.
OM 5179217–1–100