D The generator can supply a maximum of 100 W to the tube.
This example represents one 30 kV/200mAs exposure each minute.
Line Conditioner Resonance Frequency
1-1-6
D 20 kHz approx.
Generator Power (excluding tube)
1-1-7
D 5 kW maximum.
1-2
Maximum Tolerance of Displayed Constants
kV and mAs:
D kV: "5%
D mAs: "7.6% "1.1 mAs
Measurement Conditions:
D kVp:
Connect an HV voltage divider (Machlett Dynalizer IIIA; ratio 10 kV/1 V) in series with the
x-ray tube and measure the voltage with a multimeter (Fluke 47)
D mAs:
Connect an HV voltage divider (Machlett Dynalizer IIIA; ratio 10 kV/1 V) in series with the
x-ray tube. Connect a waveform analyzer (Tektronix 7854) to the output of the divider; the analyzer
receives a signal proportional to the voltage applied to the tube. The mAs values are given by:
mAs =
Where T1 is the time at which the high voltage reaches 75% of its maximum value, and
T is the time at which the high voltage returns to 75% of the maximum value.
Compression Force and Breast Thickness:
D Compression Force:
D Breast Thickness
1-3
Workstation Storage
Total internal disk capacity: two disks, each of 9 Gbytes.
Allocated for image storage: 14 Gbytes (corresponding to approximately 1650 images).
1-4
Digital Detector FOV
Field of View (FOV) of the Digital Detector: 19 cm x 23 cm.
T2
i dt
T1
"10 newton
"10 mm
Senographe 2000 D Acquisition System