GE Senographe 2000 D Operator's Manual page 148

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
E24
GEN
None
E25
GEN
None
E26
GEN
None
E35
GEN
None
E36
GEN
None
E50
GAN
Failure
Message on
Explanation
Control Console
Failure before
Current image
exposure
acquisition aborted
before exposure due
to failure in the
AWS/SENO
A problem occurs in
Synchro. fail
synchronization
between AWS system
and generator during
Next exposure is
AWS/SENO
AWS system cannot
comm. failure
communicate with
SYSTEM
A power OFF of the
SHUTDOWN
whole system has
STARTED
been started.
The generator will
power off the supply of
Detector, Gantry and
Generator in fiew
CART PWR
The system shutdown
FAILURE – SYST
has been processed
SHUTDOWN
because an UPS error
has been detected or
communication with
UPS haas been lost.
Software Failure
Gantry software error.
146
Senographe 2000 D Acquisition System
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
equipment
1. Switch OFF the whole system and
wait for the end of the shutdown
sequence
2. Restart the system
acquisition.
4. If the problem persists, note ALL
messages and inform GEMS Field
inhinited
Service.
1. Switch OFF the Gantry
2. Switch ON
Generator.
3. If the probleme occurs again at the
next acquisition power OFF the
whole system and wait for the end of
the shutdown sequence.
4. Restart the system
5. If the problem persists, note ALL
messages and inform GEMS Field
Service.
Stop system usage and wait the end
of the shutdown sequence.
minutes.
1. Stop system usage and wait the
end of the shutdown sequence.
2. Check the main power supply.
3. Check (on UPS panel) that the
batteries are charging. Do not
attempt to restart the system until the
bateries have recharged.
4.If the problem persists, note ALL
messages and inform GEMS Field
Service.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and error codes
(ESPECIALLY THE 6 DIGITS OF
E50 MESSAGES) and contact
GEMS Field Service.
OM 5179217–1–100