REV 1
Message
Msg.
on Column
Source
No.
Readout
I28
GEN
None
I29
GEN
None
I31
GEN
None
I32
GEN
None
I33
GEN
None
I34
GEN
None
I46
GEN
None
I50
GAN
Lock
Jammed
I51
GAN
Hot Lock
I52
GAN
Arm
Unlocked
I54
GAN
Wait...
Message on
Explanation
Control Console
Max wait for
It is necessary to wait
same exposure
the displayed time to
mn
be able to repeat the
same exposure.
Excessive mAs
The amount of mAs
for this mode
necessary to repeat
mAs
the same exposure is
too high in the mode
Abort: Image
signal too low
Abort: Image
signal too high
Abort: No AOP
zone found
Thickness out of
AOP range
System Setting in
AWS set-up in
progress
progress (loading of
Flat Field map) after a
new filter was
Lock Jammed
LOCK PROBLEM:
In spite of command,
lock release does not
Lock Overheated
Lock temperature too
high (used for too long
period and/or too
Tube Housing
Tube housing arm in
Arm Unlocked
unstable position
because lock is badly
Power Limitation
Heat protection for 27
V supply. Elevator,
rotation, and light
centering device lock
are blocked.
151
Senographe 2000 D Acquisition System
Wait the required amount of time to
repeat exposure. If an exposure is
attempted before this time, an abort
may possibly occur (there is a safety
margin to ensure proper operation).
You can ignore the message if a
different, less dense, breast is to be
imaged next.
Switch to a different mode or to a
different filter.
selected.
No action required.
This message is displayed for two
seconds after selection of a new
filter.
selected.
Try again to release lock.
If the problem continues, note ALL
messages and contact GEMS Field
operate.
Service.
Wait for lock to cool.
frequently)
Pivot arm to latch correctly.
If the problem continues, note ALL
messages and contact GEMS Field
latched.
Service.
Leave equipment idle for a while.
If the problem continues, note ALL
messages and contact GEMS Field
Service.