GE Senographe 2000 D Operator's Manual page 164

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message on AWS screen
BB
"AWS reset in progress:
4
please wait"
BB
"AWS reset successful,
5
System is operational"
BB
Communication with UPS not
6
established. Please call
service. System will power off
within 30 seconds: do NOT
log in now!
BB
IDC/SENO communication
7
failure, exam forbidden
"Remote connection in
BB
8
progress : do not acquire
images"
BB
Restart browser must be
9
performed to take into
account the new configuration
System currently performing
BB
10
self–testing – Please Wait.
CC These messages relate to internal disk or memory problems:
CC
"ATTENTION Not enough
1
space to continue"
CC
"AWS not available for
2
acquisition, exam forbidden"
Disk space is full, you cannot
CC
3
perform a new exam
Senographe 2000 D Acquisition System
Explanation
An AWS boot or reset sequence
is in progress.
The AWS may start an automatic
reset if it detects an internal
acquisition device error.
The AWS boot or reset sequence
has been successfully completed
This message is displayed at the
end of the power-up/boot
sequence, if communication
cannot be established with the
UPS.
Communication between the IDC
and the Gantry system was not
established at boot time, or has
been lost during application.
A GE Insite connection is in
progress; no new images should
be acquired.
Medical Application Preferences
setttings have been changed,
and Save has been requested.
The changes will not take effect
until the Browser has been
restarted.
The system is busy with internal
checks and tests.
Image disk is full.
The Medical Application cannot
allow acquisition because of lack
of disk space or an uncomplete
action on the AWS
Image disk is full.
162
OM 5179217–1–100
Corrective action
Wait until system initialization is complete.
If several automatic resets occur, contact
GEMS Field Service.
No action required (click OK to close the
pop-up window).
1. Wait for the end of the shutdown
sequence.
2. Try to restart the system .
3. If the problem is repeated, contact
GEMS Field Service.
1. Click OK on the pop-up window
2. Reset the gantry (power OFF/ON)
3. If the message is repeated, perform
system shutdown and restart the system.
4. If the problem persists, contact GEMS
Field Service.
No action is required; wait for completion
of the connection.
Restart the Browser.
No action required; wait for completion of
the test.
1. Increase free space on image disk by
deleting exams that have been archived.
2. If the problem persists, contact GEMS
Field Service.
1. Check that all actions are complete
(e.g., close all pop-up windows) and
2. Increase free space on image disk by
deleting exams that have been archived.
2. If the problem persists, contact GEMS
Field Service.
1. Increase free space on image disk by
deleting exams that have been archived.
2. If the problem persists, contact GEMS
Field Service.