GE Senographe 2000 D Operator's Manual page 157

Acquisition system
Table of Contents

Advertisement

GE Healthcare
REV 1
Message
Msg.
on Column
Source
No.
Readout
S14
GEN
None
S20
GEN
None
S21
GEN
None
S22
GEN
None
S23
GEN
None
S25
GEN
None
Message on
Explanation
Control Console
Maximum mAs
Maximum mAs set at
exceeded
installation are
exceeded. (This
message appears only
if a value below the
maximum possible
mAs value was set at
Acq. Abort, image
Exposure has been
can be lost
fully performed but
image may be lost
because an error
occurred in the image
chain detection or
Waiting for AWS
Communication with
boot
AWS system is not yet
Waiting for IDC
Communication with
boot
Image Detection
Control sysetm is not
yet established.
Laterality is not
You have not selected
selected
the laterality of the
breast to perform
acquisition. Exposure
Tube housing
The temperature of the
over temperature
tube housing is too
high (> 65°C).
155
Senographe 2000 D Acquisition System
Lower mAs values.
installation).
1. Press RESET.
2. check image quality if image is
displayed
2. Reposition the patient.
3. Start exam again
2. If the problem persists, power OFF
processing.
the whole system and wait for the
end of the shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
1. Wait AWS boot and login is
finished.
established.
2. If message persists 10mn after the
AWS login power OFF the whole
system and wait for the end of the
shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
1. Wait Image detection Control boot
is finished.
2. If message persists 10mn power
OFF the whole system and wait for
the end of the shutdown sequence.
5. Restart the system
6. If the problem persists again, note
ALL messages and inform GEMS
Field Service.
Select laterality on the Control
Console. If you are imaging a test
object, the laterality chosen is not
important.
is inhibited
Wait for the tube housing
temperature to fall.
OM 5179217–1–100