9.2 Technical Alarm Information

Message
SpO
2
SpO
Sensor Off
2
SpO
Sensor Err
2
SpO
No Sensor
2
SpO
Comm Fail
2
SpO
Low Perfusion
2
(EDAN SpO
)
2
SpO
Noisy Signal
2
(EDAN SpO
)
2
Cause
SpO
sensor
may
2
disconnected
from
patient measuring site.
Malfunction in the SpO
sensor or in the extension
cable.
No
SpO
sensor
2
connected to the monitor.
SpO
module failure or
2
communication failure
The pulse signal is too
weak or the perfusion of
the measurement site is
too low.
There is interference with
SpO
measurement
2
signals due to patient
movement, ambient light,
electrical interference or
else.
- 56 -
Alarm Level
be
the
User-selectable
2
Low
was
Low
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Low
Low
Alarm Information
Action Taken
Make
sure
the
sensor
is
well
connected to the
patient's finger or
other parts. Make
sure the monitor
and cables are well
connected.
Replace the SpO
2
sensor
or
the
extension cable.
Make
sure
the
monitor and sensor
are well connected,
reconnect
the
sensor.
Stop
using
measuring function
of SpO
module,
2
and
notify
biomedical engineer
or
manufacturer's
service staff.
Reconnect the SpO
2
sensor and change
the
measurement
site.
If
problem
exists, please notify
biomedical engineer
or
manufacturer's
service staff.
Check the condition
of patient and avoid
patient movement;
make sure the cable
is well connected.