Vital Signs Monitor User Manual
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which iM3 is used exceeds the
applicable RF compliance level above, iM3 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating iM3.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
c
The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are
6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and
40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are
1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz,
10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to
21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
Table 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
Test
Band
frequency
(MHz)
(MHz)
385
380-390
450
430-470
communications equipment
a)
a)
Service
Modulation
TETRA
400
modulation
GMRS
FM
460, FRS
460
transmitters, as determined by an
electromagnetic site survey,
less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of
following symbol:
b)
Maximum
power
(W)
Pulse
1.8
b)
18
Hz
c)
± 5 kHz
deviation
1 kHz sine
- 148 -
EMC Information
b
equipment
marked
Distance
(m)
0.3
2
0.3
a
should be
with
the
Immunity
test level
(V/m)
27
28