Vital Signs Monitor User Manual
Monitoring SpO
2

11.11.2 SatSeconds "Safety Net"

The SatSeconds "Safety Net" is for patients whose saturation makes frequent excursions below or
above the SpO
limit but does not remain in violation long enough for the SatSeconds limit to be
2
reached. If three or more SpO
alarm limit violations occur within a 60-second period, an alarm
2
will be triggered even if the SatSeconds limit has not been reached.

11.11.3 Setting SatSeconds Duration

You can set SatSeconds to Off or to the duration among 10, 25, 50 and 100. To configure the
SatSeconds settings, enter the SpO
Setup menu and select the desired SatSeconds setting from
2
the SatSeconds list.