EDAN iM3 User Manual page 69

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Message
F3000 TEMP
TEMP Error E1
TEMP Error E2
TEMP Error E3
TEMP Error E4
Probe Heater error
TEMP Error P2
TEMP Error P3
TEMP Error P4
TEMP Error P5
Cause
System
error
synchronization.
System
error
patient
thermistor
calibration.
System error during heater
thermistor calibration.
System timing error.
Heater error.
Monitor
Mode
thermistor unstable or out
of range.
Monitor
Mode
thermistor unstable or out
of range.
Predict
Mode
thermistor unstable or out
of range.
Predict
Mode
thermistor unstable or out
of range.
- 61 -
Alarm Level
during
Medium
during
Medium
Medium
Medium
Medium
patient
Low
heater
Low
patient
Low
heater
Low
Alarm Information
Action Taken
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.