Vital Signs Monitor User Manual
1
When the alarm is set to Off, the monitor won't give an alarm prompt even if an alarm
occurs. In order to avoid endangering the patient's life, the user should use this
function cautiously.
2
Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
3
Setting alarm limits to extreme values may cause the alarm system to become
ineffective. It is recommended to use the default settings.

8.3.2 Audio Alarm Paused

In monitor mode, you can temporarily prevent alarms from sounding by pressing shortcut key
.
You can set the alarm pause time as desired. The default alarm pause time is 120 s.
1. Select Menu > User Maintain, and enter the required password.
2. Select Alarm, and set Pause Time to 60 s, 120 s, or 180 s.
When alarms are paused,
The audio alarm is turned off, and no alarms are sounding.
The visual alarm indications are still displayed.
The monitor displays the audio alarm paused icon
The monitor displays the remaining pause time in seconds with red background.
When the alarm pause time expires, the audio alarm paused status is automatically terminated and
alarm is sounding. You can also terminate the alarm paused status by pressing shortcut key
.
NOTE:
If a new alarm occurs during the audio alarm paused period, the new alarm will not be
sounding.

8.3.3 Audio Alarm off

In monitor mode, set Pause Time to Permanent, press shortcut key
displays information: please confirm whether to activate audio alarm off function? Click Yes,
the monitor will enter into audio alarm off status. Click No, the monitor will keep current status.
The audio alarm is turned off, and no alarms are sounding.
The visual alarm indications are still displayed.
WARNING
.