Other operations, such as changing patient type, entering DEMO or standby mode and so on, can
stop the measurement. Meanwhile, parameter interface will be updated;
At the end of the last measurement, if user presses NIBP measurement key before monitor gets
average value with DU... sounding, a new Ward round of average measurement will be started
with prompt NIBP AVG-1 (n) and NIBP average value of the last Ward round will not be
displayed. '-1 (n)' means the first measurement of N times.
5. Stopping measurement
During measurement, press the
measurement.
NOTE:
1
In auto or average mode, when measurement is in process, pressing the
button can only stop the current measurement, and cannot end the entire
measurement mode. At the end of countdown, monitor continues to finish the NIBP
measurement; when measurement is in interval, pressing the
start measurement in advance.
2
To end the whole AVG measurement, please click AVG measurement shortcut key,
or click Stop in NIBP setup menu.

13.5.2 Correcting the Measurement if Limb is not at Heart Level

To correct the measurement if the limb is not at heart level to the displayed value:
Add 0.75 mmHg (0.10 kPa) for each centimeter
higher or
Add 1.9 mmHg (0.25 kPa) for each inch higher

13.6 NIBP Multi-Review Window

In monitor mode, clicking the triangle icon at the bottom of main interface can display/hide NIBP
multi-review window.

13.7 Resetting NIBP

Available for EDAN module only.
When the pressure does not work properly and the system fails to give a message for the problem,
pick Reset in the User Maintain > NIBP menu to activate self-test procedure, and thus restore
the system from abnormal performance.
button on the front panel at any time to stop
Deduct 0.75 mmHg (0.10 kPa) for each
centimeter lower or
Deduct 1.9 mmHg (0.25 kPa) for each inch
lower
- 85 -
Monitoring NIBP