Vital Signs Monitor User Manual
Message
SpO
2
Interference
(EDAN SpO
)
2
NIBP
NIBP Comm Fail
NIBP Leak
NIBP
Excessive
Pressure
NIBP
Init
Pressure
High
Cause
Light
Ambient light around the
sensor is too strong.
NIBP module failure or
communication failure
NIBP pump, valve, cuff or
tube has a leakage.
Pressure has exceeded the
specified
limit.
The initial pressure is too
high during measuring
Alarm Level
Low
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Low
upper
safety
Low
Low
- 57 -
Alarm Information
Action Taken
Reduce interference
of the ambient light
and avoid sensor's
exposure to strong
light.
Stop
using
measuring function
of NIBP module,
and
notify
biomedical engineer
or
manufacturer's
service staff.
Check
the
connections and the
wrapped cuff to see
whether they are all
prepared well.
If failure persists,
please
notify
biomedical
engineer
or
manufacturer's
service staff.
Measure again, if
failure persists, stop
measuring function
of NIBP module
and
notify
biomedical
engineer
or
manufacturer's
service staff.