Temp - EDAN iM3 User Manual

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
A.6 TEMP
Complies with ISO 80601-2-56: 2017+A1:2018.
T2A Module:
☆Measuring range
Working temperature
Transport and Storage
Sensor type
☆Adjustable range of alarm limits
Resolution
☆Accuracy
Response time
Update time
Warm-up time
Time for predicting
Calibration
Measuring Mode
Transient Response Time
Clinical Bias
Limits of Agreement
Stability
NOTE:
The direct mode refers to monitor mode, while adjusted mode refers to predict mode.
TH Module:
☆Measuring range
Working temperature
Transport and Storage
☆Adjustable range of alarm limits
Resolution
Response time
Monitor mode: 25 °C~45 °C
Predict mode: 35.5 °C~42 °C
10 °C ~40 °C
-20 °C ~ 55 °C
Oral /axillary /rectal
35.5 °C ~42 °C
0.1 °C
Monitor mode: ±0.1 °C (25 °C~ 45 °C)
< 60 s
1 s~ 2 s
<10 s
<30 s
Self-test interval: ≤5 min
Direct Mode/ Adjusted Mode
≤30 s
(-0.2 to -0.4 )°C
0.49
0.14 °C
34 °C~ 42.2 °C
10 °C~ 40 °C
-20 °C ~ 50 °C
35.5 °C~ 42 °C
0.1 °C
1 s
- 140 -
Product Specification

Advertisement

Table of Contents
loading

Table of Contents