Electromagnetic Immunity - EDAN iM3 User Manual

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
B.3 Electromagnetic Immunity
Guidance and manufacture's declaration – electromagnetic immunity
iM3 is intended for use in the electromagnetic environment specified below. The customer or
the user of iM3 should assure that they are used in such an environment.
IEC/EN 60601 test
Immunity test
3 V
Conducted RF
150 kHz to 80 MHz
IEC/EN
61000-4-6
6Vrms
bands between
Radiated RF
0.15 MHz and 80
MHz
IEC/EN
61000-4-3
3 V/m
80 MHz to 2.7 GHz
See table 1
level
3 V
rms
150 kHz to
80 MHz
c
in
ISM
6Vrms
ISM
between
0.15
and 80 MHz
3 V/m
80 MHz to
2.7 GHz
Comply with
table 1
Compliance
Electromagnetic environment -
level
Portable
communications equipment should be
used no closer to any part of iM3,
including
recommended
calculated
applicable to the frequency of the
transmitter.
rms
Recommended separation distance
=
d
1
c
in
bands
MHz
=
d
1
=
d
2
d=6
communications
(Portable
equipment (including peripherals such
as
antennas) should be used no closer
than 30 cm (12 inches) to any part of
the iM3, including cables specified by
the manufacturer).
Where P is the maximum output
power rating of the transmitter in watts
(W) according to the transmitter
manufacturer
recommended separation distance in
meters (m).
Field
- 147 -
EMC Information
guidance
and
mobile
cables,
separation
from
the
2 .
P
150KHz to 80MHz
2 .
P
80 MHz to 800 MHz
3 .
P
800 MHz to 2.7 GHz
P
/E
at
RF
equipment
RF
communications
antenna
cables
and
and
strengths
from
RF
than
the
distance
equation
wireless
bands
external
d
is
the
fixed
RF

Advertisement

Table of Contents
loading

Table of Contents