EDAN iM3 User Manual page 144

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
☆Adult Mode
☆Pediatric Mode
☆Neonatal Mode
☆Alarm Type
☆ Cuff
Pressure
Range
Pressure Resolution
☆Maximum Mean Error
☆Maximum Standard Deviation
Maximum Measuring Period
Adult/Pediatric
Neonate
Typical Measuring Period
Dual Independent Channel Overpressure Protection
Adult
Pediatric
Neonatal
Pre-inflation Pressure
Adult Mode
Pediatric Mode
Neonatal Mode
SYS: 40 mmHg to 270 mmHg
DIA: 10 mmHg to 215 mmHg
MAP: 20 mmHg to 235 mmHg
SYS: 40 mmHg to 230 mmHg
DIA: 10 mmHg to 180 mmHg
MAP: 20 mmHg to 195 mmHg
SYS: 40 mmHg to 135 mmHg
DIA: 10 mmHg to 100 mmHg
MAP: 20 mmHg to 110 mmHg
SYS, DIA, MAP
0 mmHg to 300 mmHg
Measuring
1 mmHg
±5 mmHg
8 mmHg
120 s
90 s
20 s to 35 s (depend on HR/motion disturbance)
(297±3) mmHg
(245±3) mmHg
(147±3) mmHg
Default: 160 mmHg
Range: 80/100/120/140/150/160/180/200/220/240 mmHg
Default: 140 mmHg
Range: 80/100/120/140/150/160/180/200 mmHg
Default: 100 mmHg
Range: 60/70/80/100/120 mmHg