13.5.1 Operation Prompts

1. Manual Measuring
Access the NIBP Setup > Function menu and set the Measure Mode item to Manual. Then
press the
button on the front panel to start a manual measurement. During measurement,
pressing this button at any time can stop measurement.
2. Automatical Measurement
Access the NIBP Setup > Function menu and set the Measure Mode item to Auto, select time
interval as need, then press the
If measurement is failed or monitor receives the command for suspending, measurement will be
stopped. The next measurement will be continued after interval. If alarms of NIBP Excessive
Pressure or NIBP Aux Excessive Pressure are present, the whole measurement will be stopped.
Manually operation is necessary for a new auto measurement.
During interval, if interval is changed, after ending of new interval countdown, monitor will start
measurement automatically.
During interval, if user manually starts the measurement, the next auto measurement will start
after the ending of countdown which starts from the latest manual measurement; During interval,
if user makes a continuous measurement, the whole measurement will be over on the basis of
continuous measurement.
3. Continuous measurement
Access the NIBP Setup menu; click Continuous to start a continuous measurement. The
continuous measurement will last 5 minutes.
4. AVG measurement
Access NIBP Setup > Function menu, set AVG Measurement interval and AVG Measurement
times, and click NIBP AVG; or click AVG shortcut key
During measurement:
Invalid value will not be used for counting average. Invalid value every measurement will result
in invalid average;
There will have sounding of DU... after entire measurement is finished. At that time, a final
average value can be displayed. Meanwhile, alert information of the last measurement and
technical alarms of the last measurement will be cleared, and alert information can work
normally. The value for each measurement and final average value will be stored in Ward round
records/spot-checking trend table which uses icon * to mark the average value;
In specified interval, if interval and times have been changed again, the whole average
measurement will be ended and measurement times will be zero;
button on the front panel. During measurement:
- 84 -
Monitoring NIBP
directly to start measurement.