Cleaning And Disinfecting Other Accessories; After Reprocessing; Storage And Transport - EDAN iM3 User Manual

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
1.
Disassemble quick TEMP module from the monitor.
2.
Clean and dry the TEMP Sensor according to the methods in section 19.3.2.3 prior to
disinfection.
3.
Prepare the disinfectant solution.
4.
Wipe the sensor with a soft cloth dampened with the disinfectant solution (ethanol or
isopropanol).
5.
Wipe off the disinfectant solution with a dry cloth after disinfection.
6.
Leave the sensor to air dry.
7.
Inspect the TEMP Sensor, to ensure that there is no damage.

19.5 Cleaning and Disinfecting Other Accessories

For cleaning and disinfecting other accessories, refer to the instructions delivered with the
accessories. If the accessories are not accompanied by instructions, refer to this manual for the
methods of cleaning and disinfecting the monitor.

19.6 After Reprocessing

After reprocessing, the equipment, cables, cuffs, sensors and other accessories
should be checked to ensure there are no signs of aging, wear, cracks, deformation,
discoloration or peeling, etc. Replacement should be taken or contact the service
personal of the manufacturer if necessary.
Assembling and attaching device-specific components
Prerequisite:
All components have been reprocessed and are dry.
Preparation before next use of device
Assembling and fitting patient-specific accessories and consumables, i.e. SpO
Cuffs.

19.7 Storage and Transport

After reprocessing, there are no special requirements for storage and transport of the product.
However, the following must be observed:
– Store dry and free of dust
– Avoid recontamination and damage during transport
All further information on storage and transport included in the accompanying documents must
be observed.
- 123 -
Care and Cleaning
sensors and NIBP
2

Advertisement

Table of Contents
loading

Table of Contents