Known Issues And Directions - EDAN iM3 User Manual

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
TEMP alarm limits are listed as follows:
Patient Type
ADU/PED

9.5 Known Issues and Directions

No.
Known issues
1
When recording the trend graph in
monitor
shortcut key is pressed to stop
recording in the process of recording
the second parameter trend graph,
the paper for recording will not be
left blank after the second parameter
trend graph.
2
When USB flash drive is inserted
via OTG interface, it may not be
recognized.
3
In
Admit/Create new patient hardkey
4
Switching Network Type may lead
to Wi-Fi unavailable.
5
In Ward round or spot-checking
mode, the Removable Device can
be selected, while storage device is
fixed as Internal Storage Device.
Adjustable Range
+35.5 °C (+95.9 ℉) ~ +42 °C (+107.6 ℉)
mode,
if
the
spot-checking
mode,
, there is no key sound.
Directions
Re-print trend graph, and do not
Record
press Record shortcut key in
printing process.
Re-plug USB flash drive
press
N/A
If Wi-Fi is unavailable, please
restart the monitor (refer to
Section Front View) under the
precondition of ensuring patient's
safety.
N/A
- 67 -
Alarm Information
Step
+0.1 °C (32.2 ℉)

Advertisement

Table of Contents
loading

Table of Contents