EDAN iM3 User Manual page 71

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Message
Others
Battery Low
Battery Error
Recorder Out Of Paper
Recorder
Setup
Needed
Insufficient
storage
space
Read-only
storage
device
Cause
Battery Low
Malfunction in Battery
Recorder Out Of Paper
The
user
presses
RECORD button when
Recorder
is
configured.
Less than 10M space is
left in the device.
The device is read-only.
- 63 -
Alarm Level
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Low
Low
the
Low
not
Low
Low
Alarm Information
Action Taken
Please change the
battery or charging.
Replace the battery
and
restart
the
monitor.
If
the
problem
persists,
notify
the
manufacturer's
service staff.
Please install the
paper.
Notify
the
manufacturer's
service
staff
to
install and set the
recorder.
Delete some data in
the device or use
another device.
Repair the device or
replace it with a
new one.

Advertisement