EDAN iM3 User Manual page 49

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
NOTE:
When the monitor is turned on for the first time or after upgrading software, please modify
User Maintain password according to the prompts. The default initial User Maintain
password is ABC.
Set CMS/Gateway Encryption to Off, AES (default) or TLS when user connects the
monitor with network server (MFM-CMS or gateway).
Click Load Certificate to install/upgrade the Certificate via USB flash drive. The certificate
issued by Certificate Authority (CA) is recommended and self-signed certificate should be
avoided.
Steps of importing certificate for MFM-CMS and Gateway:
Step 1, in CA Certificate column, select corresponding CA certificate whose suffix
should be '.cer'.
Step 2, in Client Certificate column, select corresponding client certificate whose suffix
should be '.cer'.
Step 3, in Private Key Certificate column, select corresponding private key whose suffix
should be '.pem'.
Step 4, in Password column, input corresponding password.
Step 5, click Confirm.
NOTE:
1
CA certificate is the root certificate of CA institution.
2
Client certificate/Server certificate are the SSL certificate obtained from CA institution
and it is recommended to use OV-SSL certificate type.
3
The private key shall be generated by the RSA-2048 algorithm and be encrypted by
using the AES-256 algorithm. The password for encryption shall be at least 8
characters (a combination of letters and numbers).
4
The certificate format is as follows: For CA and Client certificate/Server certificate,
'.cer' is supported; for private key, '.pem' is supported. All the certificates should
comply with X.509 format.
5
The storage path for certificate related to TLS enterprise-level encryption method:
For MFM-CMS\gateway: root directory of USB flash drive\certs\cms\
Set Firewall to On to protect against hacker attacking.
Click Firewall Rules to check rule details.
Set Packets Limit value for traffic monitoring. If the data traffic into the monitor in Per
minute exceeds the threshold, the monitor will remind the user of the exceeding and the risk
of network attack.
Set Auto Login to On/Off.
- 41 -
Networking

Advertisement

Table of Contents
loading