Vital Signs Monitor User Manual
whose hands are unavailable.
Place the finger into the sensor according to the direction of the symbol on the sensor. The
fleshiest part of digit should be covering the detector window.
Orient the sensor so that the cable will be running towards the top of the patient's hand.
Connect the sensor with the monitor (or with the extension cable if needed).
Applying Neonatal Finger (or Toe) Wrap Sensors:
When you perform the measurement, position the sensor over the hand or foot with optical
components opposite each other.
Hold the sensor, and insert stretched strap into slot, hold it there while threading end through
latch. If strap is too long, thread it through second latch.
Connect the sensor with the monitor (or with the extension cable if needed).
Applying Adult/Pediatric Ear Clip Sensor:
When you perform the measurement, clip the plastic fixing mechanism on top of the ear; make
the ear afford the sensor's weight to avoid the sensor getting loose.
Clip the probe onto fleshy part of the lobe with optical components opposite to each other.
Connect the sensor with the monitor (or with the extension cable if needed).
3. Plug the connector of the sensor extension cable into the SpO
Inspect the application site every two to three hours to ensure skin quality and correct
optical alignment. If the skin quality changes, move the sensor to another site. Change
the application site at least every four hours. For neonate, change the measuring site
every 20 minutes.
WARNING
- 72 -
Monitoring SpO
socket.
2
2