EDAN iM3 User Manual page 70

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Message
TEMP Error P6
Temp COMM Fail
Temp exceed limit
No TEMP Sensor
Measure Site Error
Probe TEMP Too High
Cause
Unable to pre-heat probe
tip.
TEMP module failure or
communication failure.
The TEMP value is out of
the range of +30 °C ~ +43
°C.
Probe configuration (or no
probe connected) error.
The probe in use is not
consistent
with
measure position set on
the monitor.
The original temperature
of sensor >+33 °C and
≤+40 °C.
- 62 -
Alarm Level
Low
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Medium
Low
the
Medium
Low
Alarm Information
Action Taken
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.
NOTE:
Measure
readings
displayed on the
screen
are
unreliable
when
the
monitor
indicates
Temp
Error P06.
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.
Put the probe into
the probe well; take
it out and measure
again.
Well connect the
probe
and
the
monitor,
and
measure again.
Correctly set the
measure position on
the monitor.
Please
wait
for
probe temp drop
and start measuring.