EDAN iM3 User Manual page 154

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Power frequency
(50/60Hz)
magnetic field
IEC/EN 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
± 8 kV contact
± 15 kV air
±2 kV for power supply
lines
±1 kV for line to line
±2 kV for line to ground
30 A/m
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles )
Single phase: at 0°
0 % UT; 250/300 cycle
± 8 kV contact
± 15 kV air
±2kV for power
supply lines
±1 kV for line to
line
±2 kV for line to
ground
30 A/m
0 % U
0.5 cycle
T;
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles )
Single phase: at 0°
0 % U
; 250/300
T
cycle
- 146 -
EMC Information
Floors should be wood,
concrete or ceramic tile.
If floor are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power
frequency
magnetic fields should
be
at
levels
characteristic
of
typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of iM3 requires
continued
operation
during
power
mains
interruptions,
it
recommended that iM3
be powered from an
uninterruptible
power
supply or a battery.
a
is