NOTE:
1. When the monitor is connected with the central monitoring system, you must set
Disconnect Alarm to On.
2. If Disconnect Alarm occurs during audio alarm paused or audio alarm off status, the
monitor will prompt a sounding alarm with information of NetWork Disconnect.
During the network disconnected status, activating audio alarm paused or audio alarm
off function can disable the audio alarm signal of Disconnect Alarm.

7.4 e-link Function

A monitor with e-link module can transmit data through e-link and also displays the connection
status between the monitor and the communication device in the general information area.
To enable/disable the function, select User Maintain > input password > select Network > set
e-link to Active Connection, Passive Connection or Off (default).
Active Connection: used to actively receive data from external module.
In System Setup > e-link, user can view Paired Devices list and Available Devices list.
Select from the Available Devices list to Connect with communication device.
In Paired Devices list, up to 10 recently connected devices can be displayed. Select one of
them to view details (such as Device Name, Alias, MAC address, etc.) or to delete. User can
also modify Alias which supports 1~20 English characters.
Passive Connection: used to send data to the network in response to HL7 requirement.
In User Maintain > Network, user can set e-link Name and e-link Key.
E-link name setup for monitors of same model has fixed rules. 20 characters are supported at
most. For example: iM3 default e-link name "VS_3".
Connection permission (e.g. pairing code) is needed for the first connection. Pairing code
supports 6 numbers at most. Default pairing code is 123456.
For Passive Connection, the transmitted data include: patient information (in monitor/Ward
round/spot-checking mode, doctor, blood type and department information are excluded),
parameter measurement data.
The monitor responds to the communication device's request to transmit data. The data to be
transmitted include:
The data displayed in current interface: NIBP, SpO
infrared TEMP data), transmission time.
NOTE:
For Passive Connection, the supported data transmission format is XML or ER7, and the
supported e-link version is BLE4.0, which is applicable in 3 system environments: iOS,
Android and Windows. For Active Connection, only version below 4.0 is supported.
, TEMP (prediction data of Quick TEMP and