EDAN iM3 User Manual page 66

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Message
NIBP Aux Excessive
Pressure
NIBP Time Out
NIBP Self Test Error
NIBP Cuff Type Error
(EDAN NIBP)
NIBP
SystemPress.Abnormal
NIBP System Failure
NIBP Weak Signal
Cause
Pressure has exceeded the
second safety limit as
specified.
Measuring
time
exceeded the specified
time.
Sensor or other hardware
errors.
The cuff type used isn't
consistent with the patient
type.
Atmospheric pressure or
system
pressure
abnormal. The valve is
occluded so that deflation
is failed.
NIBP is not calibrated.
Cuff is too loose or
patient pulse is too weak.
- 58 -
Alarm Level
High
has
Low
High
Low
is
Low
High
Low
Alarm Information
Action Taken
Notify biomedical
engineer
or
manufacturer's
service staff.
Measure again or
use other measuring
method.
Measure again, if
failure persists, stop
using
measuring
function of NIBP
module and notify
biomedical
engineer
or
manufacturer's
service staff.
Confirm the patient
type and change the
cuff.
Check whether the
airway is occluded
or pressure sensor
works properly. If
the problem still
exists, contact your
service personnel.
Contact
your
service personnel.
Use other methods
to measure blood
pressure.