Vital Signs Monitor User Manual
Message
NIBP Range Exceeded
NIBP Loose Cuff
NIBP Interference
T2A TEMP
Temp Comm Fail
Temp Exceed Limit
No TEMP Sensor
Ambient TEMP Too
High
Ambient TEMP Too
Low
Cause
Maybe the patient blood
pressure value is beyond
the measurement range.
Cuff
is
not
properly
wrapped or no cuff is
connected.
Signal noise is too large
or pulse rate is not regular
due
to
the
movement.
TEMP module failure or
communication failure.
The
TEMP
value
beyond the range of 35.5
°C ~ 42 °C
TEMP
sensor
is
connected to the TEMP
module.
The Sensor temperature is
higher than +40 °C
The Sensor temperature is
lower than +10 °C
- 59 -
Alarm Level
High
Low
Low
patient
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
is
Medium
not
Low
Low
Low
Alarm Information
Action Taken
Use other methods
to measure blood
pressure.
Properly wrap the
cuff.
Make sure that the
patient
under
monitoring
is
motionless.
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.
Put the sensor into
the sensor bracket,
take
it
out
and
measure again.
Connect the sensor
and
the
monitor
well, and measure
again.
Put the sensor into
the sensor bracket,
measure again after
the
ambient
temperature reaches
normal value.