Injected dyes such as methylene blue or intravascular dyshemoglobins such as
methemoglobin and carboxyhemoglobin may lead to inaccurate measurements.
2
Inspect the sensor to ensure that the light emitter and receiver are aligned with each
other and there is no gap between the sensor and the finger. All the light emitted by
the light emitter must pass through the patient's tissue. The sensor cable should be
placed on the back of the hand.
3
Clean and remove any substances such as nail polish from the application site.
Periodically check to ensure that the sensor remains properly positioned on the
patient.

11.4 Measurement Limitations

Certain patient conditions can affect the measurements or cause the loss of the pulse signal.
Inaccurate measurements can be caused but not limited by:
incorrect sensor application
high levels of ambient light sources, such as surgical lights (especially those with a xenon
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight
failure to cover the sensor with opaque material in high levels of ambient light conditions
dysfunctional hemoglobins
low peripheral perfusion
excessive or violent patient movement
venous pulsations
intravascular dyes, such as indocyanine green or methylene blue
externally applied coloring agents (nail polish, dye, pigmented cream)
defibrillation
placement of the sensor on an extremity with a blood pressure cuff, arterial catheter, or
intravascular line
electromagnetic interference
Loss of pulse signal can occur for the following reasons:
the sensor is applied too tightly
a blood pressure cuff is inflated on the same extremity as the one with the sensor attached
there is arterial occlusion proximal to the sensor
low peripheral perfusion