EDAN iM3 User Manual page 68

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Message
Probe Is Missing
Probe TEMP Too High
Probe Heater Error
Temp SENSOR OFF
Measure Site Error
TH TEMP
Temp exceed limit
Cause
The
probe
connected
cannot be identified.
The original temperature
of sensor >+33 °C and
≤+40 °C.
Heater error.
The probe is disconnected
from the patient.
The probe in use is not
consistent
with
measure position set on
the monitor.
The
TEMP
value
beyond the range of 34 °C
~ 42.2 °C
- 60 -
Alarm Level
Medium
Low
Medium
Low
the
Medium
is
Medium
Alarm Information
Action Taken
Please change probe
or
contact
Manufacturer's
service staff.
Please
wait
for
probe temp drop
and start measuring.
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.
Connect the sensor
and the patient well,
and measure again.
Correctly set the
measure position on
the monitor.
Check the integrity
of the probe cover,
make sure it is
clean, and take a
new measurement.