54 . Lambda 265, 365 and 465 ASX-260/280 AND ASX-520/560 Autosampler Users Guide
Scan and Wavelength Program Mode Measurement
Procedure
Auto Sampler accessory is available to be controlled in Scan, Wavelength Program,
NOTE:
Quantification and Scanning Quantification mode of UV Express software.
Start the System Self Test after more than 20 minutes warming-up of the system.
NOTE:
1.
Double-click UV Express
modes for starting.
2.
The following window will be shown. Empty the cell holder and close the lid firmly.
When performing System Self Test, both reference and sample cell holder have to be
NOTE:
emptied. Please remove cuvettes including the flow cell from cell holder.
3.
Click OK after finishing the System Self Test.
4.
Click New to open a new window. Select "Experiment Type" and click OK.
5.
Click Method for set up parameter and then click Accessory tab. For more detail of
method refer to UV Express Software Users Guide.
6.
Select the Auto Sampler accessory in the Accessory tab and then Click Method
Setup.
folder and select Scan (or Wavelength Program)