PANNORAMIC® SCAN RX 1.0
Exposure Settings
You can set the Exposure time weight (%) here (displayed under TW in the Channel Settings
frame). Set Lumencor excitation intensity at the Excitation intensity (%) field. The defined
value will be used as maximum when applying Auto Exposition time if you activate the Max.
excit. intensity option.
Caution!
To prevent over-exposure, the Excitation intensity (%) value can be modified by the
software when calculating Auto Exposure time.
Image Intensity
Click
to open Image intensity panel. An image is defined as a two-dimensional function
f(x, y) the amplitude of at any pair of coordinates (x, y) is called the intensity or gray level of
the image at that point. The pixel intensity at the coordinate pointing at with the cursor is
displayed in the Image Intensity window.
5.2.2.4 Stage Positioning
After clicking
control function. The number at the center of the controller defines the step size by which
the stage is moved. Value can be entered by hand or with the slider.
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Figure 84. Window of Image Intensity
you can perform microscope movements on the Live image with this
Figure 85. Stage positioning panel
SCANNING MODES AND SETTINGS