PANNORAMIC® SCAN RX 1.0
Scan settings
Change filters after whole slide scanning
To optimize speed, filter change should take place after the digitization of the slide with
one channel.
Change filters at each field-of-view
Filters should be changed for each field of view if very precise colocalization is required
but consider that this option is more time consuming.
Notice:
Make sure to select the Change filters at each field of view option if using multi band filter
for multi-channel digitization.
Caution!
If the specimen area is barely visible, it is possible that the Pannoramic® Scanner Software
for Research control software will analyze that area as an empty one.
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Figure 67. Focus level and step size
Figure 68. Scan settings
SCANNING MODES AND SETTINGS
84