PANNORAMIC® SCAN RX 1.0
▪
Barcode area light
▪
Backlight illumination intensity
o
Objective guide
▪
Step size
▪
X / Y values or navigation arrows
▪
Goto
▪
Goto Preview Origin
o
RGB LED
▪
Set illumination intensity per channel
o
Sync with microscope
o
Calculate white balance
o
Save RGB intensity
•
–
Slide load tester
Slide loading operation can be checked for each of the slides in a magazine. Click Insert
magazine if the prepared magazine or magazines are placed at the bottom of the input
guide rails. Click Start slide insert test button to launch the process. Slides are loaded in
but preview image will not be recorded.
If Magazine counter is set to -1, all the magazines will be tested in cycles based on the
value set in the Loading cycle per magazine field.
Click Remove magazine when finished.
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Figure 126. Slide load tester
SETTINGS AND SERVICES