2.4 Coupling the Lumencor SOLA SM Unit with the Scanner

To operate the Lumencor SOLA SM fluorescent illumination unit, it has to be connected to
the Pannoramic® SCAN RX base unit.
1.
Turn the control computer and the Pannoramic® SCAN RX unit off.
2.
Connect both ends of the fiber optic cable to the coupling ports of the scanner and the
illumination unit (see Figure 2 – option A in chapter
3.
Connect the power supply cables of both the scanner and the illumination unit.

2.5 Coupling the Lumencor Spectra 6 Unit with the Scanner

To operate the Lumencor Spectra 6 Light Engine fluorescent illumination unit, it has to be
connected to the Pannoramic® SCAN RX base unit, the camera, and the control computer.
1.
Turn the control computer and the pco.edge 4.2 bi camera off.
2.
Attach Lumencor SPECTRA 6 fluorescent illumination unit to the coupling adapter of the
Pannoramic® SCAN RX scanner (see Figure 2 – option B in chapter
3.
Connect power supply cables of both the scanner and the illumination unit.
4.
Connect TTL trigger cable between pco.edge 4.2 bi camera (out 4) and the Lumencor
SPECTRA 6 unit (TTL port), and USB-to-RS232 communication cable between the
Lumencor SPECTRA 6 unit and the computer.
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .