PANNORAMIC® SCAN RX 1.0
Caution!
It requires lots of time (since it cannot find focus neither on the edges, nor on specks) to
digitize coverslip edges. Select Remove coverslip function to eliminate those areas.
Warning!
Do not use Remove coverslip function on TMA slides! The layout of TMA slides may mislead
the detecting algorithm.
Fill holes in scanmap
If selected, the holes inside the specimen will also be scanned. Often these holes contain
small sample objects that otherwise would not be scanned.
Remove specks
If selected, the scanner does not digitize objects smaller than the predetermined size. This
helps filtering specks and other uninteresting objects out.
Set the maximum volume of specks manually in the setting window
Threshold
Auto threshold
Select this option to automatically detect the specimen in the preview image.
Caution!
Make sure that the detection result is acceptable for you. Otherwise set the threshold value
manually.