3D Histech PANNORAMIC SCAN RX User Manual page 39

Table of Contents

Advertisement

PANNORAMIC® SCAN RX 1.0
Unload the slide by clicking Unload next to slide image and unload the magazine by
clicking Unload all next to the rack image. Set magazine number in the value box at the
left of the control bar, then click Change to load another magazine in the feeder.
If there is a lens marking at the bottom of the position preview in the magazine image,
the preview image of that slide has already been created, and if a question mark is
displayed the slide was not yet loaded. A slide that is loaded at the moment is highlighted
with a blue background and outer line.
Objective changer
Live view is on
Live view is on with the shutter closed
Click
to display the toolbar for Live view. These functions are only available for the displayed
live image (i.e., Live view is ON), but not available for camera images during digitization.
When in Normal live mode, images are taken with the user-specified exposure time.
Notice:
Enter full-screen live mode by pressing the F11 key. For more information, see section
Hotkeys.
One pixel on the monitor corresponds to one pixel on the camera.
The live image completely fills the display area.
Displays or hides the vertical and horizontal scale bars
Displays the magnifier window
Activates red marking on overexposed pixels
Saves the current image as BMP or JPEG file
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Figure 13. Slide/tray control
Toolbar for live images
SOFTWARE DESCRIPTION
4.3