PANNORAMIC® SCAN RX 1.0
o
Set Position, Objective type, and Magnification settings of the objective.
o
Use manual camera changer – Changing the cameras is performed manually.
o
Pre-scan objective protection enabled – If activated, slide area near the edges will
not be scanned to record fast preview image. If inactive, Fast Preview processing
becomes slower.
o
Cameras – Activate BF and/or FL mode for the camera. The software automatically
loads the camera settings upon clicking the Test and apply camera settings
button.
Preset profiles are the following:
▪
Calibrated Linear: Color scheme is defined by calibration (Gamma value is
not included).
▪
sRGB standard: Color scheme is defined by calibration (standard Gamma
value -2.2- is applied).
▪
sRGB standard Microscope Like: Color scheme is defined by calibration
(standard Gamma value / White Point: D50).
▪
Legacy: A Color scheme calibrated for the previous software version is
applied.
▪
Custom: parameters are freely customizable as follows:
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
▪
Color Scheme
▪
Calibrated: Scheme is defined by the color calibration slide.
▪
Legacy: A Color scheme calibrated for the previous
software version is applied.
▪
Raw: the color information of the recorded raw image is
applied.
▪
White Point
▪
D65: color temperature according to a standard illuminant
D65 white point value is applied (6500K / daylight).
▪
D50: color temperature according to a standard illuminant
D50 white point value is applied (5000K / horizontal
daylight).
▪
Gamma
▪
Standard: a value of 2.2 is used.
▪
Medium: a value of 1.8 is used.
▪
Off: a value of 1.0 is used.
SETTINGS AND SERVICES
126