5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Magazine jammed
1. Exit control software.
2. Switch off control computer and 24V power supply, disconnect power cable.
3. Remove protective covers (see chapter
4. Remove magazine manually. If a jammed magazine cannot be removed, contact
service personnel.
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Virtual slide is improperly focused almost across its entire surface.
Focus range enabled
Disable Limiting focus range.
Poor image quality
Imaging error (for example, chromatic aberrations)
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory settings give the most
accurate images.
Virtual slides show poor stitching in Viewer.
Scan camera is not properly adjusted (for example, due to loose fixture)
Contact 3DHISTECH Service for support.