PANNORAMIC® SCAN RX 1.0
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
•
Magazine jammed
1. Exit control software.
2. Switch off control computer and 24V power supply, disconnect power cable.
3. Remove protective covers (see chapter
4. Remove magazine manually. If a jammed magazine cannot be removed, contact
service personnel.
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Virtual slide is improperly focused almost across its entire surface.
•
Focus range enabled
•
Disable Limiting focus range.
Poor image quality
•
Imaging error (for example, chromatic aberrations)
•
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory settings give the most
accurate images.
Virtual slides show poor stitching in Viewer.
•
Scan camera is not properly adjusted (for example, due to loose fixture)
•
Contact 3DHISTECH Service for support.
MAY 13, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
TROUBLESHOOTING AND MAINTENANCE
9.5.1 Removing the Protective
Covers).
146