Dräger 8000 IC Instructions For Use Manual page 50

Table of Contents

Advertisement

What's What
Skin temperature control module
3.1 On / off button for skin temperature control.
3.2 Calibration stamp (only in Germany).
3.3 Key to display peripheral skin temperature
(only when ThermoMonitoring option is installed).
3.4 Green Control LED;
is lit when skin temperature control is switched on.
3.5 Display for actual value (measured value)
of skin temperature.
3.6 Display for set value of skin temperature
left key: to decrease set value
right key: to increase set value.
3.7 Red ±0.5 °C LED;
flashes/is lit when the actual value of skin
temperature deviates from the set value by more
than 0.5 °C.
3.8 Red Sensor LED;
flashes/is lit when the skin temperature sensor is
faulty or disconnected.
3.9 Certification mark.
3.10 Key for checking measurement electronics
Display 36 ± 0.1 °C (reference).
50
°C HAUT/SKIN/PEAU/PIEL
3.5
°C
°C
3.6
3.7
± 0,5 °C
Check
15.22
36 °C
87.06
±0,1
Sensor
3.9
3.10
3.8
3.4
Control
3.3
3.2
3.1
Control

Advertisement