Operation

Controlling air temperature

Controlling air temperature
Measure the patient's core temperature at regular
intervals.
Fold down flap with brief instructions.
1 Push on / off switch in – until it engages = on.
2 Green LED is lit.
Wait for self-test to finish, then incubator will switch to
"air temperature control" mode.
3 The display alternates regularly between measured
value for air temperature and the word, SEt.
4 Default value for flashing display of air temperature
set value is 33.0 °C.
5 Press P or p key briefly to select this set value.
The display will remain continuously lit.
Setting values in 28 °C to 37 °C range
The set value can be changed in 0.1 °C steps.
6 Press p key – the set value is decreased,
7 Press P key – the set value is increased.
Press P or p key until the set value required is
reached.