Using Electronic Baby Scales; Using Vacuum Mattress - Dräger 8000 IC Instructions For Use Manual

Table of Contents

Advertisement

Using electronic baby scales

Accessory
1 Set bed at maximum height with handwheels.
Open front door.
2 Slide the electronic scales under bed as far as
possible. Guide scales along ridges on base plate.
Rail carrying display unit passes through the lower
hose seal.
Close front door again and lower bed to its lowest
setting.
Weigh baby, following Instructions for Use for Seca
Inscale Baby Scales.
After weighing:
Take scales out in reverse order
or
Scales may remain in incubator.
Bed can be tilted, as before.

Using vacuum mattress

Accessory
The mattress can be formed into any shape, and will
keep this shape after the mattress has been evacuated.
In this way babies can be supported in extreme positions
for special treatment. The standard mattress may also
remain inside the incubator.
Open front door.
Place mattress inside and form it into approximately
the shape required.
Place baby on mattress and "mould" mattress around
baby.
Connect hose of suction unit to vacuum mattress.
Open valve and evacuate vacuum mattress.
Close valve and remove hose.
Close front door.
1
Operation
Baby scales
Vacuum mattress
1
2
35