Air temperature control module
2.1 Key to switch on air temperature control (only when
skin temperature control module is fitted).
2.2 Unlocking key for setting a higher range
of values up to 39 °C.
2.3 Yellow >37 °C LED;
is lit when a higher range of values has been set.
R R R R
2.4 Green
LED;
indicates heating-up phase.
2.5 Green Control LED;
is lit when air temperature control is switched on
(only when skin temperature control module is
installed).
2.6 Display for actual value (measured value) of
air temperature.
2.7 Display for set value of air temperature;
left key: to decrease set value
right key: to increase set value.
2.8 Red ±1.5 °C LED;
flashes/is lit when the actual value of
the air temperature deviates from the set
value by more than 1.5 °C.
2.9 Red Sensor LED;
flashes when air temperature sensor is faulty.
S S S S
2.10 Red
LED;
flashes for fan failure.
T T T T
2.11 Red
LED for high temperature;
flashes / is lit when air temperature is or was higher
than 38 °C (40 °C for higher range of set values)
and in advance of Inop-alarm.
2.12 Key to reset high temperature alarm.
°C LUFT/AIR
2.6
°C
°C
2.7
2.8
±1,5 °C
Reset
Sensor
2.9
2.10
2.11
2.12
What's What
2.5
Control
2.4
2.3
>37 °C
2.2
>37 °C
2.1
Control
49