Dräger 8000 IC Instructions For Use Manual page 43

Table of Contents

Advertisement

Skin temperature control module (optional)
Fault
Red ±0.5 °C LED flashes.
Actual value display flashes.
Intermittent tone commences.
Red Sensor LED flashes.
Three dashes flash in centre of
actual value display.
Continuous tone commences.
Err flashes in actual value display.
Red Inop. LED is lit in main module.
Continuous tone commences.
Err is lit in actual value display.
Reference temperature of 36 °C
is outside tolerance of ±0.1 °C
when 6 key is pressed.
Three dashes are lit at the top in
actual value display.
Three dashes are lit at bottom of
actual value display.
Cause
Skin temperature deviating from
set value by more than ±0.5 °C.
Skin temperature sensor not
properly connected.
Faulty skin temperature sensor.
Fault in skin temperature control
module while in operation.
Fault in the skin temperature
control module; skin temperature
control switched off.
Temperature measuring unit not
operating correctly.
Temperature measured above 42 °C.
Temperature measured below 30 °C.
Fault, Cause, Remedy
Skin temperature control module
°C HAUT/SKIN/PEAU/PIEL
Control
°C
°C
± 0,5 °C
Check
15.22
Control
36 °C
87.06
±0,1
Sensor
Remedy
Below set value:
check that sensor is properly fitted
to patient.
Above set value:
measure core temperature of patient
and inform doctor responsible for
patient immediately.
Check connection and put right, if
necessary.
Replace sensor.
Switch to air temperature control.
Call DrägerService.
Other modules continue to operate
properly.
Call DrägerService.
Call DrägerService.
Cooling required.
Inform doctor responsible for
patient immediately.
Check that sensor is properly fitted
to patient.
43